MEGGER
®
BMM2000 Series
Insulation Multimeters
User Guide
SAFETY WARNINGS
Safety Warnings and Precautions must be read and
understood before the instrument is used. They
must be
observed during use.
The circuit under test must be de-energised and isolated
before connections are made except for voltage measurement.
Circuit connections must not be touched during a test.
After insulation tests, capacitive circuits must be allowed to
discharge
before disconnecting the test leads.
The Live Circuit Warning and Automatic Discharge are additional
safety features and
should not be regarded as a substitute for
normal safe working practice.
Replacement fuses must be of the correct type and rating.
Failure to fit the correctly rated fuse
will result in damage to the
instrument in the event of an overload.
Test leads, including crocodile clips, must be in good order,
clean and have no broken or cracked insulation.
Ensure that hands remain behind guards of probes/clips when
testing.
U.K. Safety Authorities recommend the use of fused test leads
when measuring voltage on high energy systems.
NOTE
THE INSTRUMENTS MUST ONLY BE USED BY
SUITABLY TRAINED AND COMPETENT PERSONS
Users of this equipment and/or their employers are reminded that
Health and Safety Legislation require them to carry out valid risk
assessments of all electrical work so as to identify potential
sources of electrical danger and risk of electrical injury such as
from inadvertent short circuits. Where the assessments show that
the risk is significant then the use of fused test leads constructed
in accordance with the HSE guidance note GS38 ‘Electrical Test
Equipment for use by Electricians’ should be used.
Symbols used on the instruments are:
Caution, risk of electric shock.
Caution, refer to User Guide.
Equipment protected throughout by
Double Insulation (Class II).
Equipment complies with current EU Directives.
Equipment must not be connected
to installations >500V.
>500V
Seitenansicht 0
1 2 3 4 5 6 ... 13 14

Inhaltsverzeichnis

Seite 1 - User Guide

MEGGER®BMM2000 SeriesInsulation Multimeters User GuideSAFETY WARNINGS• Safety Warnings and Precautions must be read andunderstood before the instrumen

Seite 2 - OPERATION

Zero offset at probe tips: 0,10Ω typicalLead resistance zeroing: Up to 9,99ΩNoise rejection: 1V rms 50/60HzBuzzer: Operates at less than 5Ω (approx

Seite 3

then calculate the measurement range. This is the range ofmeasurement over which the error in service is less than 30%of the reading. Digital instrume

Seite 4 - 10V Range

Complies with the following parts of EN61557, Electrical safety inlow voltage systems up to 1000V a.c. and 1500V d.c. –Equipment for testing, measurin

Seite 5

REPAIR AND WARRANTYThe instrument circuit contains static sensitive devices, andcare must be taken in handling the printed circuit board. If theprotec

Seite 6

AVO INTERNATIONAL Archcliffe RoadDoverKent CT17 9EN Tel: +44 (0) 1304 502100England Fax: +44 (0) 1304 2073424271 Bronze WayDallas Tel: +1 (800) 723-

Seite 7

GENERAL DESCRIPTIONThe MEGGER BMM2000 Series instruments are batterypowered Insulation and Continuity testers, with a measurementcapability from 0,01Ω

Seite 8

Live Circuit WarningWhen more than 25V (10V on 10V insulation range) is appliedto the terminals in the insulation ranges, the instrument defaultsto a

Seite 9 - SPECIFICATION

Note: There is a short delayon the first operation of‘1000V’ range, eachtime the range isselected. This is toprevent accidentalapplication of 1kV.The

Seite 10

Continuity Testing (Ω) (See fig.2)The continuity tests are activatedwhen the probes make contact of lessthan a few kΩ. The tests apply aconstant curre

Seite 11

Display: Audible:<5Ω continuous bleep<3kΩ short bleep>3kΩ no bleepResistance Tests (kΩ)This is a low voltage (5V) low current (25µA) test for

Seite 12

Zeroing of d.c. mV (no a.c. mV zero facility)To zero the d.c. mV range, short the leads together in thed.c. mV position, wait for the reading to settl

Seite 13

1. Set the selector switch to mA.2. Connect the test leads.3. Press and hold down the TEST button for approximately 2 seconds.4. Select either ac or

Seite 14 - AVO INTERNATIONAL

SPECIFICATION(All quoted accuracies are at +20°C.)Insulation RangesNominal Test Voltage (d.c.):BMM2080: 50V, 100V, 250V, 500V, 1000VBMM2000: 250V, 500