Megger BMM2000 Series Betriebsanweisung Seite 9

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 14
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 8
SPECIFICATION
(All quoted accuracies are at +20°C.)
Insulation Ranges
Nominal Test Voltage (d.c.):
BMM2080: 50V, 100V, 250V, 500V, 1000V
BMM2000: 250V, 500V, 1000V
BMM2000ESD: 10V, 100V, 500V
Test voltage accuracy: 1000V, 500V, 250V, 100V, 50V,
+15% maximum on open circuit.
10V, ±10% maximum on open
circuit.
Short circuit current: < 2 mA
Test Current on load: 1mA at min. pass value of
insulation specified in BS7671,
HD384 and IEC364, 2mA max.
Accuracy (BMM2080)
Range Full Scale Accuracy
1000V 200G ±2% ±2digits ±0,2% per G
500V 100G ±2% ±2 digits ±0,4% per G
250V 50G ±2% ±2 digits ±0,8% per G
100V 20G ±2% ±2 digits ±2,0% per G
50V 10G ±2% ±2 digits±4,0% per G
Accuracy (BMM2000)
Range Full Scale Accuracy
1000V 20G ±2% ±2digits ±0,2% per G
500V 10G ±2% ±2 digits ±0,4% per G
250V 5G ±2% ±2 digits ±0,8% per G
Accuracy (BMM2000ESD)
Range Full Scale Accuracy
500V 10G ±2% ±2digits ±0,4% per G
100V 2G ±2% ±2 digits ±2,0% per G
10V 1G ±2% ±2 digits ±2,0% per 100M
Note: Above specifications only apply when high quality silicone
leads are being used.
Measuring Range: 0,01Mto 200G
(0-100Gon analogue scale).
EN61557 Operating range: 0,10Mto 1,00G
Leakage Current: 10% ±3digits
Continuity
Measuring Range: 0,01to 99,9
(0-10on analogue scale)
EN61557 Operating range: 0,10to 99,9
Accuracy: ±2% ±2 digits
Open circuit voltage: 5V ±1V
Test current: 210mA ±10mA (0-2)
Seitenansicht 8